
18. Troubleshooting
18-8 HomeChoice APD Systems Patient At-Home Guide October 2, 2009 07-19-61-244
To Correct
Cause 2:
If you suspect this is the cause:
1. Press
STOP.
2. Press V until you see MANUAL DRAIN.
3. Press
ENTER to initiate a Manual Drain. Then open the
clamp on the patient line.
Do not open the clamp until after you press
ENTER.
4. When the Manual Drain ends, STOPPED: FILL appears.
5. Press
GO to resume Fill.
Cause 3: CHECK HEATER LINE appears at start of FILL 1 only.
The pump chambers are not able to fill from the heater or
empty to the drain completely.
Blocked heater or drain line.
To Correct
Cause 3:
1. Check heater line.
2. Check drain line.
h NOTE: Even though CHECK HEATER LINE appears on
the display screen, it is possible that the cause is
a blocked drain line.
CHECK LINE (Continued)
Komentáře k této Příručce