07-19-61-244 October 2, 2009 HomeChoice APD Systems Patient At-Home Guide 18-15
18. Troubleshooting
18.3.7 Low Drain Volume
7. Press ENTER to bypass.
END OF THERAPY or LAST FILL appears.
– or –
WARNING
Bypassing a LOW DRAIN VOLUME alarm can leave fluid in
the peritoneal cavity and result in an increased
intraperitoneal volume (IIPV) situation.
IIPV can result in a feeling of abdominal discomfort,
serious injury, or death.
If any patient, or patient caregiver, suspects the patient has
IIPV during a treatment, press
STOP immediately, then
press
V and initiate a Manual Drain. The Manual Drain
procedure is located in 18.5, Manual Drain Procedure, on
page 18-53. See 18.8, Increased Intraperitoneal Volume
(IIPV), on page 18-58 if IIPV is suspected.
Additional care should be taken to monitor for IIPV
symptoms for those patients not able to communicate
essential information to their caregiver during treatment.
Steps to bypass LOW UF alarm (Continued) Display screen
END OF THERAPY
LAST FILL
Zobrazit stránku 269